SuperNova v428: New Bright Polymer Dye for Flow Cytometry
Arunkumar Easwaran, Massimiliano Tomasulo, Boi Hoa San, Mukesh Mudgal, Sergei Gulnik | Beckman Coulter Inc., Miami, FL, USA
Valérie Mallet Designe, Sylvain Monseaux | Beckman Coulter, Marseille, France
Introduction to SuperNova v428
SuperNova v428 is a new violet laser (405 nm)-excitable water-soluble polymer dye which shows exceptional brightness. This is attributed to its high extinction coefficient (~3 million Lmol-1 cm-1) and fluorescence quantum yield (~0.6) in phosphate buffer. Excitation maximum of this polymer is measured as 414 nm and its emission maximum is at 428 nm (Figure 1) which can be detected using a 450/50 bandpass filter or equivalent in a flow cytometer.
Figure 1: (Left) Absorption and emission spectra (λexc = 405 nm) of SuperNova v428 in PBS. (Right) SuperNova v428 solution in PBS upon illumination using a UV lamp at 365 nm. |
Synthesis of SuperNova Polymer dyes and conjugation to Antibodies
SuperNova v428 dye is prepared by polymerization of water-soluble monomers as shown in Figure 2 which leads to formation of a highly conjugated fluorescent backbone. Capping is carried out on the polymer followed by activation using appropriate functionalities, resulting in polymer capable of conjugation to antibodies. These activated polymers conjugated to antibodies followed by purification using a standardized procedure.
Figure 2: Schematic procedure for SuperNova v428 polymer synthesis, activation and conjugation process. |
Polymer dye conjugates usually tend to generate significant non-specific staining, for instance by binding nonspecifically to cells expressing Fc receptors such as monocytes. A proprietary formulation has been identified and optimized to minimize this non-specific staining, allowing to further improve stain index of SuperNova v428.
Figure 3: CD22-SNv428 whole blood staining before (left) and after (right) formulation with additives preventing non specific staining. Background on monocytes is significantly reduced. |
Flow Cytometry results
SuperNova v428 was conjugated to dierent antibodies and compared with PE, Pacific Blue and Brilliant Violet polymer dye conjugates. Antibody-SuperNova v428 conjugates exhibited more than 10 times the brightness of Pacific Blue and also better brightness than PE and BV 421 in flow cytometry. Nonspecific staining was also significantly reduced compared to BV 421 conjugates.
Figure 4: SuperNova v428 conjugates (with CD19, CD22, and CD25) compared against conjugates of BV421, PE and Pacific Blue in flow cytometry. All SuperNova v428 conjugates showed higher brightness to above mentioned commercially available conjugates. Nonspecific binding significantly reduced in the case of SuperNova v428 conjugates. |
Lymphocytes Stain Index |
Monocytes MFI (background) |
||
---|---|---|---|
CD19 | SN v428 | 281 | 0.688 |
BV421 | 191 | 1.533 | |
PE | 246 | 0.805 | |
Pacific Blue | 48 | 0.600 | |
CD22 | SN v428 | 172 | 1.033 |
BV421 | 123 | 1.476 | |
PE | 136 | 0.716 | |
Pacific Blue | 28 | 0.958 | |
CD25 | SN v428 | 36 | 0.731 |
BV421 | 14 | 2.021 | |
PE | 24 | 1.052 | |
Pacific Blue | 6 | 0.522 |
Table 1: Lymphocytes staining index and monocytes MFI (background) obtained staining normal whole blood with lymphocytes markers conjugated to SN v428, BV421, PE and Pacific Blue. |
Conclusion
SuperNova polymer dyes are a new generation of polymer dyes for flow cytometry application. SuperNova v428 has unique photo-physical properties leading to extremely bright conjugates when conjugated to antibodies. A proprietary formulation leveraging unique additives allows to minimize non-specific staining for greater confidence in flow cytometry staining results and improved staining index. Therefore SuperNova v428 conjugates are particularly suited for the study of low abundance and dimly expressed markers.
For Research Use Only – Not for use in diagnostic procedures.
Pacific Blue is a registered trademark of Molecular Probes, Inc.
Helpful links
-
Flow Cytometry Reagents
-
Antibodies & Kits
-
Single-Color Antibodies
- Isotype Controls
- Panel Design
- Product Listings and Catalogs
- Quality Standards
- APO2.7 (7A6) Antibodies
- Beta2-Microglobulin Antibodies
- CD10 Antibodies
- CD103 Antibodies
- CD105 Antibodies
- CD106 Antibodies
- CD107a (LAMP-1) Antibodies
- CD109 (E123) Antibodies
- CD116 (GM-CSFR) Antibodies
- CD117 (c-kit) Antibodies
- CD11a Antibodies
- CD11b Antibodies
- CD11c Antibodies
- CD120b (TNF-R2) Antibodies
- CD122 (IL-2Rß) Antibodies
- CD123 (IL-3Rα) Antibodies
- CD126 (IL-6Rα) Antibodies
- CD127 (IL-7Rα) Antibodies
- CD13 Antibodies
- CD133 Antibodies
- CD134 (OX 40) Antibodies
- CD135 (flt3/flk2) Antibodies
- CD137 Antibodies
- CD138 Antibodies
- CD141 (BDCA-3) Antibodies
- CD14 Antibodies
- CD144 (VE-Cadherin) Antibodies
- CD146 Antibodies
- CD15 Antibodies
- CD152 (CTLA-4) Antibodies
- CD154 Antibodies
- CD157 Antibodies
- CD158a,h Antibodies
- CD158b1/b2,j Antibodies
- CD158e1/e2 Antibodies
- CD158f (KIR2DL5) Antibodies
- CD158i Antibodies
- CD159a Antibodies
- CD16 Antibodies
- CD160 Antibodies
- CD161 (NKR-P1A) Antibodies
- CD166 (ALCAM) Antibodies
- CD16b Antibodies
- CD18 Antibodies
- CD180 Antibodies
- CD183 (CXCR3) Antibodies
- CD184 (CXCR4) Antibodies
- CD19 Antibodies
- CD196 (CCR6) Antibodies
- CD197 (CCR7) Antibodies
- CD1a Antibodies
- CD1c Antibodies
- CD2 Antibodies
- CD20 Antibodies
- CD200 Antibodies
- CD203c Antibodies
- CD206 (MMR) Antibodies
- CD207 (Langerin) Antibodies
- CD208 (DC-LAMP) Antibodies
- CD209 (DC-SIGN) Antibodies
- CD21 Antibodies
- CD22 Antibodies
- CD223 Antibodies
- CD23 Antibodies
- CD235a (Glycophorin A) Antibodies
- CD24 Antibodies
- CD243 (P-glycoprotein) Antibodies
- CD244 (2B4) Antibodies
- CD25 Antibodies
- CD26 Antibodies
- CD27 Antibodies
- CD274 (PD-L1) Antibodies
- CD278 (ICOS) Antibodies
- CD279 (PD-1) Antibodies
- CD28 Antibodies
- CD29 Antibodies
- CD294 (CRTH2) Antibodies
- CD3 Antibodies
- CD30 Antibodies
- CD300a (IRp60) Antibodies
- CD305 (LAIR1) Antibodies
- CD309 (KDR) Antibodies
- CD31 Antibodies
- CD314 (NKG2D) Antibodies
- CD32 Antibodies
- CD326 (EpCAM) Antibodies
- CD328 (p75 AIRM1) Antibodies
- CD33 Antibodies
- CD335 (NKp46) Antibodies
- CD336 (NKp44) Antibodies
- CD337 (NKp30) Antibodies
- CD34 Antibodies
- CD35 Antibodies
- CD352 (NTB-A) Antibodies
- CD36 Antibodies
- CD366 (TIM-3) Antibodies
- CD37 Antibodies
- CD370 (Clec9A) Antibodies
- CD38 Antibodies
- CD39 Antibodies
- CD4 Antibodies
- CD40 Antibodies
- CD41 Antibodies
- CD42a Antibodies
- CD42b Antibodies
- CD43 Antibodies
- CD44 Antibodies
- CD45 Antibodies
- CD45RA Antibodies
- CD45RO Antibodies
- CD48 Antibodies
- CD49b Antibodies
- CD49d Antibodies
- CD49e Antibodies
- CD5 Antibodies
- CD52 Antibodies
- CD54 Antibodies
- CD55 Antibodies
- CD56 Antibodies
- CD57 Antibodies
- CD58 Antibodies
- CD59 Antibodies
- CD6 Antibodies
- CD61 Antibodies
- CD62L Antibodies
- CD62P Antibodies
- CD63 Antibodies
- CD64 Antibodies
- CD65 Antibodies
- CD66b Antibodies
- CD66c (KOR-SA3544) Antibodies
- CD69 Antibodies
- CD7 Antibodies
- CD71 Antibodies
- CD73 Antibodies
- CD77 Antibodies
- CD79a Antibodies
- CD79b Antibodies
- CD8 Antibodies
- CD80 Antibodies
- CD81 Antibodies
- CD83 Antibodies
- CD85d (ILT4) Antibodies
- CD85j (ILT2) Antibodies
- CD85k (ILT3) Antibodies
- CD86 Antibodies
- CD8β Antibodies
- CD9 Antibodies
- CD90 Antibodies
- CD94 Antibodies
- CD95 Antibodies
- CD99 Antibodies
- Cyclin A2 Antibodies
- Cytokeratin Antibodies
- DAP12 (DNAX / KARAP / TYROBP) Antibodies
- FMC7 (CD20) Antibodies
- FoxP3 Antibodies
- Granzyme B Antibodies
- Helios Antibodies
- HIV-1 Core Antibodies
- HLA-ABC Antibodies
- HLA-DQ Antibodies
- HLA-DR Antibodies
- HLA-DR, DP, DQ Antibodies
- IFN-γ Antibodies
- IgD Antibodies
- IgE (Dε2) Antibodies
- IgM Antibodies
- IL-17A Antibodies
- IL-2 Antibodies
- IL-4 Antibodies
- Kappa Chain Antibodies
- Ki-67 Antibodies
- Lactoferrin Antibodies
- Lambda Chain Antibodies
- Myeloperoxidase Antibodies
- NG2 (7.1) Antibodies
- NKp80 Antibodies
- OSCAR Antibodies
- Perforin Antibodies
- ROR1 Antibodies
- TCR PAN α/β Antibodies
- TCR PAN γ/δ Antibodies
- TCR Vα7.2 Antibodies
- TCR Vα24 Antibodies
- TCR Vβ1 Antibodies
- TCR Vβ2 Antibodies
- TCR Vβ3 Antibodies
- TCR Vβ4 Antibodies
- TCR Vβ5.1 Antibodies
- TCR Vβ5.2 Antibodies
- TCR Vβ5.3 Antibodies
- TCR Vβ7.1 Antibodies
- TCR Vβ7.2 Antibodies
- TCR Vβ8 Antibodies
- TCR Vβ9 Antibodies
- TCR Vβ11 Antibodies
- TCR Vβ12 Antibodies
- TCR Vβ13.1 Antibodies
- TCR Vβ13.2 Antibodies
- TCR Vβ13.6 Antibodies
- TCR Vβ14 Antibodies
- TCR Vβ16 Antibodies
- TCR Vβ17 Antibodies
- TCR Vβ18 Antibodies
- TCR Vβ20 Antibodies
- TCR Vβ21.3 Antibodies
- TCR Vβ22 Antibodies
- TCR Vβ23 Antibodies
- TCR Vδ1 Antibodies
- TCR Vδ2 Antibodies
- TCR Vγ9 Antibodies
- TCR ζ Antibodies
- TdT Antibodies
- Thrombin Receptor Antibodies
- TIA-1 Antibodies
- TNF Alpha Antibodies
- ZAP-70 Antibodies
- Liquid Multi-Color Antibody Cocktails
- AQUIOS Reagents
- ClearLLab Antibody Solutions
-
DURAClone Panels
- DURAClone Ecosystem
- Immune System Research
- Immune Function Research
- Rare Event Detection
- Adult Stem Cells
- DURAClone IM Count
- DURAClone IM Phenotyping Basic
- DURAClone IM T Cell Subsets
- DURAClone IM Treg
- DURAClone IM TCRs
- DURAClone IM B Cell
- DURAClone IM Dendritic Cell
- DURAClone IM Granulocytes
- DURAClone IM Innate Lymphoid Cell Tube
- DURAClone IF T Cell Activation
- DURAClone IF T Helper Cell
- DURAClone IF Monocyte Activation
- DURAClone IF Basophil Activation
- DURAClone RE CLB
- DURAClone RE PC
- DURAClone RE ALB
- DURAClone SC Mesenchymal Stem Cell
- DURAClone SC Hematopoietic Stem Cell
- DURActive Stimulation Kits
- Clinical/Clinical Research Systems & Kits
-
Single-Color Antibodies
-
Cell Health and Research Assays
- Cell Cycle Flow Cytometry Assay Principle
- Streamlined Workflow for Cell Cycle Analysis
- Cell Cycle Staining Protocol
- Advanced Cell Cycle Analysis
- Propidium Iodide Spectrum
- Cell Cycle Kit Performance
-
ViaKrome Fixable Viability Dyes
- Comparison of NHS ester-based reagents to ViaKrome Fixable Viability Dye
- Flexibility for Multicolor Panel Design
- Robust Performance
- ViaKrome Fixable Viability Dye Staining Protocol
- Why Use a Viability Dye?
- Comparison_NEW
- Flexibility for Multicolor Panel Design_NEW
- Performance_NEW
- Staining Protocol_NEW
- Why_NEW
- DNA Binding Dyes
- Apoptosis Reagents
- Research Assays
- SuperNova Fluorescent Polymer Dyes
- Lysing, Fixative, and Permeabilizating Reagents
- Fluorospheres & Quality Control
- Custom Reagent Services
-
Antibodies & Kits