Experience simplified high-throughput sample processing and more control with the JXN-26 three-phase centrifuge with powerful high speed features, low heat output, imbalance tolerance and quiet performance. Has a maximum capacity of 6.0L and RPM of 26,000 (82,000 x g). The Friction Reduction System (FRS) means the machine runs quicker, quieter and cooler, with shorter cycle times with high-torque because of the exclusive Switched Reluctance (SR) drive. Three-phase filtration breaks material into solids, liquids and gases. Stores up-to 1000 user-defined programs. This build of the JXN 26 is in vitro diagnostic (IVD) compatible.

Specifications

Platform Avanti
BioSafe Yes
Acceleration 11 acceleration profiles
Deceleration 12 profiles
Speed Control ±|10.00|rpm of set speed or 0.1%, whichever is greater
Timed Run Up to 99 hrs, 59 mins
Max Capacity Volume Range 1.00|6.00|L
Software Virtual Network Computing (VNC), Mobile Fuge app for Apple iOS and Android
Temperature Range(s) 10°C - 40°C in 1 degree increments
Rotor ID Automatic
Zonal Capability Yes
Installation Category II
Environmental Conditions -10°C to 40°C in 1° increments
Humidity Restrictions 80% (non-condensing)
Material(s) Steel, Plastic
Finish Powder Coat Paint
Weight 290.00|kg / 639.340560336145|lb
Width 71.00|cm / 27.9527559055118|in
Depth 86.00|cm / 33.8582677165354|in
Height 86.00|cm (149 cm with door open)
Clearances 7.6 cm (sides) & 16 cm (rear)
Pollution Degree 2
Power Requirements 220 V, 50 Hz

Technical Documents

NOT ALL PRODUCTS ARE AVAILABLE IN ALL COUNTRIES
PRODUCT AVAILABILITY AND REGULATORY STATUS DEPENDS ON COUNTRY REGISTRATION PER APPLICABLE REGULATIONS

The listed regulatory status for products correspond to one of the below:
IVD: In Vitro Diagnostic Products. These products are labeled "For In Vitro Diagnostic Use."
ASR: Analyte Specific Reagents. These reagents are labeled "Analyte Specific Reagents. Analytical and performance characteristics are not established."
CE: Products intended for in vitro diagnostic use and conforming to European Directive (98/79/EC). (Note: Devices may be CE marked to other directives than (98/79/EC)
RUO: Research Use Only. These products are labeled "For Research Use Only. Not for use in diagnostic procedures."
LUO: Laboratory Use Only. These products are labeled "For Laboratory Use Only."
No Regulatory Status: Non-Medical Device or non-regulated articles. Not for use in diagnostic or therapeutic procedures.