Can Sapphire windows be used for low-UV AUC experiments?

spINSIGHTS logo

img5

When do I use Quartz vs. Sapphire Windows in an Analytical Ultracentrifuge (AUC) experiment?

AUC experiments in Absorbance mode traditionally uses Quartz windows because of their excellent transmittance in the low UV range. Experiments in Interference mode use Sapphire windows because of their mechanical strength.

Sapphire vs. Quartz

img2

Sapphire / Quartz Optics are comparable

Rubicon ES2 Sapphire (red trace) and Quartz (blue trace) Intensity scans for air, buffer and BSA protein samples in panel A, B & C respectively. (D) Absorbance profile for BSA protein for both window types. (E&F) show transmittance profiles of both window types. The optical performance of Sapphire windows is comparable to Quartz down to ~ 230 nm (peptide bond peak). 

 

 

AUC performance is also comparable*

img4

Analysis of results of SV-AUC experiments at 25 krpm on BSA protein sample are shown for Sapphire (red trace) and Quartz (blue trace) at three different wavelengths – 230, 240 and 280 nm in Panels A, B & C respectively. Panel D shows Interference results at 50 krpm. The sedimentation coefficient population distributions overlap, indicating identical results obtained from both window materials.

Better Interference data with Sapphire windows*

table1

Rubicon ES2 Sapphire windows offer superior mechanical strength in comparison to Quartz windows – which results in less deformation at high speeds (> 50 k). This does not affect Absorbance data but can affect the quality of Interference data. This table shows the signal-to-noise ratio where the degradation for quartz-interference data can be seen.

Summary

New generation Rubicon ES2 Sapphire windows can be used in Absorbance mode in AUC at wavelengths as low as 230 nm, yielding performance comparable to Quartz windows. For high speed applications (e.g., peptide analysis) – they also offer less noisy Interference data vs. Quartz. 

*Analysis software has not been verified or validated by Beckman Coulter Life Sciences
For research use only. Not for use in diagnostic procedures.