PAT700 Features

The ANATEL PAT700 TOC analyzer fully complies with all global pharmacopeias (USP, EP, JP) requirements. Each sample is fully oxidized guaranteeing the most accurate TOC result that follows EP2.2.44 guidelines. Production uptime is assured with the PAT700 main and standby dual UV lamp design, making it the ideal choice for on-line TOC release. Save time and money. The PAT700 can be installed online and run up to 4 grab samples offline.

BENEFITS

Have confidence in the data - assured with UV Detect, excursion alarming, and fully traceable calibration

Implement the FDA PAT initiative for on-line release for TOC and Conductivity eliminating time consuming and costly laboratory analysis.

Reduce the risk of product recall and speed up water loop troubleshooting when an excursion alarm occurs

The PAT700 internally captures two samples and then re-analyzes the first sample to validate the excursion alarm and saves the second sample for further root-cause analysis.

Confirm production vessels are clean and ready to go on-line faster with Clean-in-Place (CIP) analysis

The PAT700 clean-in-place analysis maximizes production up-time by providing immediate on-line results for both TOC and conductivity. Providing instant on-line verification that your production vessel is ready for production changeover.

SPECIFICATIONS

 

Spec

Value

TOC

Operating Range

0.5 to 2,000 ppb as Carbon

 

Display Resolution

0.1 ppd

 

Accuracy

+/-1 ppd or +/-5% whichever is greater

 

Repeatability

+/-0.3 ppd or +/-1% whichever is greater

 

Limit of Detection

0.5 ppb

 

Maximum Input Conductivity

0.2 uS/cm for all waters, 1.0 uS/cm for all neutral waters, 5.0 uS/cm for water with CO2 as the sole conductive species

Conductivity

Conductivity

Range 0.05 to 150 uS/cm (@25* C)

 

Display Resolution

0.01 uS/cm

 

Conductivity Accuracy

+/-2% over full range (uncompensated)

 

Resistivity

Resistivity Range 0.2 to 18 MQ-cm (@25*C)

 

Display Resolution

0.01 over full range

 

Available Modes

Temperature compensated to 25*C, or uncompensated

Temperature

Ambient Operating Range

10 to 40*C (50 to 104*F)

 

Measurement Accuracy

+/-0.4*C

 

Sample Water Range

5 to 95*C (41 to 203*F)

 

Display Resolution

0.1 over full range

Physical Specs

UV Lamps

2, with UV Detect technology

 

Interface/Display

Color with touch screen

 

Maximum Altitude

4,000 m (13,125)

 

User I/O Wiring

Three, 3/4-inch conduit openings or quick disconnect fittings

 

Standards System

Onboard, Automated Standards Introduction System (OASIS)

 

Dimensions

59.7 w X 22.9 d X 25.4 h (23.5 X 9 X 10 inches)

 

Weight

13.6 kg (30 lbs)

 

Sample Inlet Flow Rate Range

60 mL/min to 300 mL/min

 

Sample Inlet Pressure Range

10 to 100 psi (69 to 690 kPa)

Compliance

Installation Category

II

 

Pollution Degree

2, IEC 61010-1

 

CE Compliance

EN 61010-1 and EN 61326

 

Safety Rating

ETL, conforming to UL 61010-1 and CSA 22.2 No. 61010-1

 

Enclosure Rating

Conduit version: IP56 Quick connect version: IP46

 

Release Tests

USP <643>, USP <645>, JP 16, EP 2.2.44

New Features

CIP

Selectable mode for Clean-In-Place analysis

 

Dual Stream option

Toggle or programmable stream switching

 

Excursion sampling

Minimum flow rate to fill excursion bottle = 160 mL/min

 

Rouge detection

Identifies oxidation cell contamination from roughing