PAT700 Features
The ANATEL PAT700 TOC analyzer fully complies with all global pharmacopeias (USP, EP, JP) requirements. Each sample is fully oxidized guaranteeing the most accurate TOC result that follows EP2.2.44 guidelines. Production uptime is assured with the PAT700 main and standby dual UV lamp design, making it the ideal choice for on-line TOC release. Save time and money. The PAT700 can be installed online and run up to 4 grab samples offline.
BENEFITS
Have confidence in the data - assured with UV Detect, excursion alarming, and fully traceable calibrationImplement the FDA PAT initiative for on-line release for TOC and Conductivity eliminating time consuming and costly laboratory analysis.
Reduce the risk of product recall and speed up water loop troubleshooting when an excursion alarm occursThe PAT700 internally captures two samples and then re-analyzes the first sample to validate the excursion alarm and saves the second sample for further root-cause analysis.
Confirm production vessels are clean and ready to go on-line faster with Clean-in-Place (CIP) analysisThe PAT700 clean-in-place analysis maximizes production up-time by providing immediate on-line results for both TOC and conductivity. Providing instant on-line verification that your production vessel is ready for production changeover.
SPECIFICATIONS
|
Spec |
Value |
TOC |
Operating Range |
0.5 to 2,000 ppb as Carbon |
|
Display Resolution |
0.1 ppd |
|
Accuracy |
+/-1 ppd or +/-5% whichever is greater |
|
Repeatability |
+/-0.3 ppd or +/-1% whichever is greater |
|
Limit of Detection |
0.5 ppb |
|
Maximum Input Conductivity |
0.2 uS/cm for all waters, 1.0 uS/cm for all neutral waters, 5.0 uS/cm for water with CO2 as the sole conductive species |
Conductivity |
Conductivity |
Range 0.05 to 150 uS/cm (@25* C) |
|
Display Resolution |
0.01 uS/cm |
|
Conductivity Accuracy |
+/-2% over full range (uncompensated) |
|
Resistivity |
Resistivity Range 0.2 to 18 MQ-cm (@25*C) |
|
Display Resolution |
0.01 over full range |
|
Available Modes |
Temperature compensated to 25*C, or uncompensated |
Temperature |
Ambient Operating Range |
10 to 40*C (50 to 104*F) |
|
Measurement Accuracy |
+/-0.4*C |
|
Sample Water Range |
5 to 95*C (41 to 203*F) |
|
Display Resolution |
0.1 over full range |
Physical Specs |
UV Lamps |
2, with UV Detect technology |
|
Interface/Display |
Color with touch screen |
|
Maximum Altitude |
4,000 m (13,125) |
|
User I/O Wiring |
Three, 3/4-inch conduit openings or quick disconnect fittings |
|
Standards System |
Onboard, Automated Standards Introduction System (OASIS) |
|
Dimensions |
59.7 w X 22.9 d X 25.4 h (23.5 X 9 X 10 inches) |
|
Weight |
13.6 kg (30 lbs) |
|
Sample Inlet Flow Rate Range |
60 mL/min to 300 mL/min |
|
Sample Inlet Pressure Range |
10 to 100 psi (69 to 690 kPa) |
Compliance |
Installation Category |
II |
|
Pollution Degree |
2, IEC 61010-1 |
|
CE Compliance |
EN 61010-1 and EN 61326 |
|
Safety Rating |
ETL, conforming to UL 61010-1 and CSA 22.2 No. 61010-1 |
|
Enclosure Rating |
Conduit version: IP56 Quick connect version: IP46 |
|
Release Tests |
USP <643>, USP <645>, JP 16, EP 2.2.44 |
New Features |
CIP |
Selectable mode for Clean-In-Place analysis |
|
Dual Stream option |
Toggle or programmable stream switching |
|
Excursion sampling |
Minimum flow rate to fill excursion bottle = 160 mL/min |
|
Rouge detection |
Identifies oxidation cell contamination from roughing |