The PAT700 TOC analyzer fully complies with all global pharmacopeias (USP, EP, JP) requirements. Each sample is fully oxidized guaranteeing the most accurate TOC result that follows EP2.2.44 guidelines. Production uptime is assured with the PAT700 main and standby dual UV lamp design, making it the ideal choice for on-line TOC release. Save time and money - the PAT700 can be installed online and run up to 4 grab samples offline.
Resources + Tools
Product Specifications
Interface Protocols | 3 x 4-20mA outputs │ 2 isolated digital inputs │ 4 x isolated digital outputs | Ethernet |
Platform | Anatel |
Industry Standards | 21 CFR Part 11, EP <2.2.44>, USP <643> |
Materials | US/Japan power cord |
Calibration Intervals and Options | 3-point TOC calibration; Conductivity calibration to USP<645>; System suitability to USP and EP |
Fluid Compatibility | Purified water, Water for injection (WFI) |
Online Operation | Sample inlet pressure 10 to 100 psi (0.69 to 6.89 Bar) |
Detection | Differential conductivity |
Reporting | ppb TOC │ µS/cm |
IP Rating | IP46 |
Installation Category | II |
Temperature Range | 5°to 95°C sample range |
Power Requirements | 100/230VAC ± 10%, 65W, 50/60Hz |
Resolution | 0.1 ppb |
Display | Full-color LCD touchscreen |
Operating Modes | Online TOC; Online Conductivity; Auto-switching Main and Standby UV lamps for 12 months operation │ 4-bottle auto-sampler for grab sample analysis │ Excursion capture for root cause analysis support |
Weight | 13.6 kg(30 lb) |
Width | 59.7 cm(23.5 in) |
Depth | 22.9 cm(9 in) |
Height | 25.4 cm(10 in) |
Color | Blue |
Package Quantity | 1 |