Particle testing in cleanroom high-pressure gas lines to ISO 14644 made easy with the MET ONE 3400 gas calibrations



This paper provides guidelines for the testing of high pressure gases used in cleanrooms according to the cleanliness limits derived from the ISO 14644-1 cleanroom classification standard, the FDA cGMP guidelines and the EU GMP Guidelines. Specifically, this applications paper relates to satisfying the air particle concentration limits of Class 100 (ISO 5) where product is exposed to the air environment.

Introduction

This paper provides guidelines for the testing of high pressure gases used in cleanrooms according to the cleanliness limits derived from the ISO 14644-1 cleanroom classification standard, the FDA cGMP guidelines and the EU GMP Guidelines. Specifically, this applications paper relates to satisfying the air particle concentration limits of Class 100 (ISO 5) where product is exposed to the air environment.

The biopharmaceutical industry routinely makes use of compressed gas in manufacturing operations. While no one standard, best practice or guideline has been specifically enforced, regulatory bodies such as the FDA take interest in the quality of compressed gas coming into contact with pharmaceutical product. Manufacturers of biopharmaceuticals have taken various approaches to the routine monitoring of compressed gas, typically based upon a risk assessment with cleanliness limits derived from the ISO 14644-1 cleanroom classification standard.

cGMP FDA Guidance to industry



The scope of particle concentration limits in the FDA guidance is not just confined to the ambient air. Compressed gases such as air, nitrogen, and carbon dioxide are routinely used in cleanrooms and are frequently employed in purging or overlaying. The FDA guidance indicates that these compressed gases should be of purity “equal to or better than that of the air in the environment into which the gas is introduced.” Section IV of the cGMP, Buildings and Facilities, addresses the need for appropriate air quality controls for aseptic processing areas of biopharmaceutical cleanrooms1. Such areas are defined as being critical because particles can enter a product as an extraneous contaminant in addition to presenting a biological contamination risk by acting as a vehicle for microorganisms.

Clean area classification (0.5µm particles/ft3 ISO designationb  0.5 particles/m3  Microbiological active air action levelsc cfu/m3  Microbiological settling plates action levelsc,d (diam. 90mm; cfu/4 hours)
100 3,520  1 1
1000  35,200 
10,000  352,000  10 
100,000  3,520,000  100  50 

Table 1. Particle limits as defined in the FDA Guidance for Industry cGMP 2004

 

EU GMP Annex 1 is less specific than the FDA Guidance. Section 81 of the guide, which covers the subject of materials required to be passed into the clean area where aseptic work takes place, indicates that non-combustible gases should be passed through micro-organism retentive filters2. Section 60 indicates that gas filters (among other critical systems and components) should be subject to validation and planned maintenance and that their return to use should be approved. Section 33 of the EU GMP, which addresses the handling and filling of aseptically prepared products, specifies that such activities should be performed in a Grade A environment with a Grade B background.

 

  Maximum permitted number of particles per m3 equal to or greater than the tabulated size 
  At rest In operation
Grade  0.5µm 5.0µm 0.5µm 5.0µm
A 3 520 20 3 520 20
B 3 520 29 352 000 2 900
C 352 000 2 900 3 520 000 29 000
D 3 520 000 29 000 Not defined Not defined

Table 2.  Particle limits as defined in EU Guidelines to Good Manufacturing Practice, Annex 1 2008

 

The High Pressure Diffuser (HPD)

 

Particle concentrations in compressed gases are typically measured using conventional air particle counters routinely used in cleanroom environmental monitoring schedules. However, direct connection of such instruments to high pressure gas lines is generally not possible. A device called a high pressure diffuser or HPD is required to reduce the pressure of the gas to be sampled to ambient such that the gas may be sampled at the instrument’s standard flow rate while at the same time not introducing particulate contamination to the gas sample. In addition to the required pressure reduction, the HPD maintains isokinetic flow through to the device output to ensure that the size-distribution of any particles suspended in the gas is not skewed.

 

The MET ONE High Pressure Diffuser or HPD in cross section indicating the flow path to the Isokinetic sampling inlet and the exhaust

Figure 1. The MET ONE High Pressure Diffuser or HPD in cross section indicating the flow path to the Isokinetic sampling inlet and the exhaust

 

The MET ONE HPD connected to a MET ONE 3400 series counter and a gas line

Figure 2. The MET ONE HPD connected to a MET ONE 3400 series counter and a gas line

 

Methodology

To sample high pressure gas lines, the HPD is installed between the connection to the gas source and the inlet of the particle counter. The presence of the HPD results in a drop in line pressure by diverting excess gas away from the sampling path and thus avoiding over pressurizing the particle counter’s sensor. Excess gas is diverted through a port in the HPD chamber either directly into the environment or away to a different location via tubing.

It is important that the HPD is supplied with gas within the specified operating range. Running the HPD at too high a pressure will result in inaccurate particle counts whereas operating below the lower limit will result in air being drawn back through the exhaust port thus contaminating the gas sample.

A second requisite for compressed gas particle counting applications is the necessity to adjust the particle counter’s sample flow control system to compensate for the density of the selected gas. Sampling at the true mass flow rate is necessary for accurate volumetric sampling to be achieved in accordance with ISO 14644 and to comply with the ISO 21501-4 calibration standard for air particle counters. The MET ONE 3400 series makes this necessary adjustment very easy and intuitive for the operator.

 

MET ONE Model 3400 series gas selection option

In this model the gas option, together with the appropriate High Pressure Diffuser (HPD), makes gas monitoring straight-forward. The user simply selects the required gas from a pull-down list on the 3400 interface. With the appropriate gas to be sampled selected, the pump control system is automatically adjusted to ensure that a 1 CFM flow rate is maintained for the selected gas.

Selecting the gas to be monitored via a simple drop-down on the MET ONE 3400 user interface

Figure 3. Selecting the gas to be monitored via a sample drop-down on the MET ONE 3400 user interface

 

Gas Sampling Procedure

 

  1. Power on the Met ONE 3400 particle counter.
  2. Program the required number and size of samples.
  3. Set the Count Mode to Counts per Cubic meter
  4. Program the concentration alarm limits per either Grade A or ISO 5 (note that the per cubic meter limits for 0.5 um are identical for both but that Grade A also specifies limits at 5.0µm
  5. Attach tubing to the exhaust port of the HPD to minimize contamination to the surrounding environment.
  6. Confirm that gas pressure at the sample point is within the HPD manufacturer’s specification (detailed in the user manual) using a calibrated pressure sensor.
  7. Fully open the sample point and purge the HPD for approximately 30 seconds.
  8. Connect the HPD to the MET ONE 3400 sample inlet nozzle using the supplied tubing.
  9. Allow gas to flow through the HPD and into the MET ONE 3400 Particle counter.
  10. Allow the system to run for one minute as a purge of the complete assembly.
  11. Initiate sampling by pressing the run button on the MET ONE 3400
  12. Once the sampling recipe has completed, close the valve on the compressed gas source.

 

Setting the sampling parameters in the MET ONE 3400

Figure 4. Setting the sampling parameters

 

Setting the alarm limits per ISO 14644 in the MET ONE 3400

Figure 5. Setting the alarm limits per ISO 14644

 

Reporting the Results

Particle results are typically reported as an average of a number of sample runs. The following example of a particle measurement on a compressed Nitrogen line demonstrates the MET ONE 3400 series support for traditional paper printout, PDF and Excel compatible files.

 

Instrument printout of a compressed Nitrogen test based upon 3 samples and an average MET ONE 3400   PDF outputs are also available on the MET ONE 3400 series portable particle counters

Excel outputs are also available on the MET ONE 3400 series portable particle counters

Figure 6,7,8. Instrument printout of a compressed Nitrogen test based upon 3 samples and an average. PDF and Excel outputs are also available on the MET ONE 3400 series portable particle counters.

 

Selecting a Particle Counter for Gas Sampling

To conduct particle testing of compressed gas lines with the MET ONE 3400 series is straight forward. Ordering a new 3400 with gas calibration can be done with a single part number. Six versions in the series are available which include gas calibrations for air, Nitrogen (N2), Carbon Dioxide (CO2) and Argon (Ar).

 

MET ONE 3400 portable w/gas calibrations PN  Description 
2088900-06 MET ONE 3413 0.3 micron, 1 CFM 
2088900-07  MET ONE 3413 0.3 micron, 1 CFM with WIFI 
2088900-08  MET ONE 3415 0.5 micron, 1 CFM 
2088900-09  MET ONE 3415 0.5 micron, 1 CFM with WIFI 

 

In addition to the MET ONE 3400 particle counter, a high pressure diffuser is required.

High pressure diffuser Description
2080732-13 1 CFM HPD 3/8" Barb

Users with existing MET ONE 3400 series particle counters wishing add gas calibration to their instruments may order this as a factory service option.

 

Calibration Description
CAL_GAS_3400 Gas calibrations for 3400 series particle counters

 

Conclusion

The MET ONE 3400 series with gas calibration option, used in conjunction with the MET ONE 3400 High Pressure Diffuser, make testing of high purity compressed gas to ISO 14644-1 easy, eliminating the potential for sampling errors associated with flow rate, thus ensuring compliance to the ISO 21501-1 calibration standard.

References

  1. Guidance for Industry. Sterile Drug Products Produced by Aseptic Processing — Current Good Manufacturing Practice. Food and Drug Administration, 2004.
  2. EU Good Manufacturing Practice, Annex 1: Manufacture of Sterile Medicinal Products

 

Watch the Webinar

Particle testing in cleanroom high-pressure gas lines to ISO 14644

 

Changes to ISO-14644-1 and -2: impact on classification methods and monitoring strategies