Cell Line Development – Limiting Dilution

Summary

  • Cell line development requires production lines start from a single cell.
  • We automated the cell counting, staining, dilution, and plating for a limiting dilution approach to achieve monclonality. This utilized:
    • Integrated Vi-CELL XR Cell Viability Analyzer for automated cell counts
    • Large volume mixing with 1200 μL capacity 96-channel head
    • Integrated plate centrifugation for immediate imaging to confirm monoclonality

 

The aspect of cell line development that arguably adds the greatest challenge to the workflow is the need to derive cell lines from a single cell. This requirement drives the well throughput into the thousands when searching for the optimal expression clone. It also adds significant time to the process as cells require weeks to grow up to usable levels and cell viability is often reduced when plating at the single-cell level. There is also the challenge of isolating a single cell per well and there are a number of different routes to achieve this.

Limiting dilution is a traditional approach to achieve monoclonality. It entails diluting cells to the level that there is one cell on average per unit of volume plated in the wells. However, as predicted by the Poisson distribution, roughly one third of the plated wells will have zero cells, one third of wells will have a single cell, and one third will have multiple cells per well (Coller and Coller 1986). This inefficiency has driven the use of alternative technologies such as single cell printers or cell sorters which can deliver single cells in nearly all wells. However, the high cost of these cell sorting devices continues to drive the use of limiting dilution In addition, cells are often exposed to the non-sterile environments prior to introduction to the cell sorter, potentially leading to contamination, while the sorting process itself (pressure, time outside incubator, etc.) can impact cell viability.

Here we describe the automation of limiting dilution on a Biomek i7 Workstation (Figure 1) that overcomes a number of challenges associated with this workflow. Plate throughput was enabled through the use of storage devices such as labware feeders and the ability to integrate numerous devices can support automated analyses. The system used a HEPA-filtered enclosure to maintain a clean environment for cell manipulations and sterile pipette tips to further ensure sample sterility. This is of the utmost importance for this workflow as the cultures will be maintained for weeks, allowing sufficient time for any contamination to spread.

Figure 1. The Biomek i7 Workstation, along with various integrated devices, was used to automate a cell line development workflow.

Figure 1. Biomek i7 Workstation with integrated devices to enable cell line development/cellular applications.

Figure 2. 1200μL-capacity multichannel head enables mixing large volumes for limiting dilution.

Figure 2. 1200μL-capacity multichannel head enables mixing large volumes for limiting dilution.

To create single cells colonies, the Biomek Workstation mixed and dispensed CHO cell suspensions to an integrated Vi-CELL XR Cell Viability Analyzer for counting. Based on the resulting cell count, ~83,300 cells were then transferred into 1mL 0.25 μM calcein AM and incubated at 37°C for 30 min. Following an initial 25-fold dilution, ~467 cells were added to a 140mL XP CHO Growth A media (Molecular Devices) such that there was one cell per 300 μL on average. Robust mixing and rapid filling of four 96-well plates was achieved through the ability to aspirate and dispense 96mL at one time with the 1200 μL-capacity multichannel head (Figure 2). Well A1 was then removed from the first plate and replaced with 300 μL of the intermediary dilution with 100X more cells to act as a control to optimize imaging focus and exposure time. Plates were spun in an integrated microplate centrifuge (Agilent) at 300 X g for 3 min to ensure any cells were at the plate bottom for immediate imaging on the CloneSelect Imager (Figure 3A, Molecular Devices). By eliminating the time for all cells to settle, we minimized the likelihood of early cell division and/or calcein stain efflux leading to false negatives.

Figure 3. Monoclonality confirmation. A) A CloneSelect Imager was used to count fluorescent cells. B) A representative plate indicating empty wells (white), monoclonal wells (green/yellow) or wells with multiple cells (red). Yellow wells indicate confirmation is required due to faint signal or abnormal shapes which can indicate cell clusters.

Figure 3. Monoclonality confirmation. A) A CloneSelect Imager was used to count fluorescent cells. B) A representative plate indicating empty wells (white), monoclonal wells (green/yellow) or wells with multiple cells (red). Yellow wells indicate confirmation is required due to faint signal or abnormal shapes which can indicate cell clusters.

Figure 3B shows the imaging results of a typical plate following automated limiting dilution plating. Monoclonal wells (green/yellow) are distinguished from those wells with zero (white) or multiple cells (red). To determine the sufficiency of mixing in the large-volume reservoir, we analyzed the frequency of monoclonality in each plate location across five experiments (20 wells).

Figure 4A shows the resulting distribution and no discernable location bias is detected. Inter-experiment consistency is demonstrated in Figure 4B, where the average frequency of monoclonal wells across the five experiments shows a narrow distribution (CV = 8.6%), at or near the expected frequency for a limiting dilution approach.

Figure 4. Consistent dilution and plating. A) Distribution of monoclonal wells across all five experiments with four plates each (20 wells per position). The conditional formatting highlights even mixing and distribution of cells in the large volume reservoir. B) Average monoclonality over four plates for each of 5 experiments illustrating consistency between experiments. The percentage of monoclonal wells approaches the expected distribution from limiting dilution.

Figure 4. Consistent dilution and plating. A) Distribution of monoclonal wells across all five experiments with four plates each (20 wells per position). The conditional formatting highlights even mixing and distribution of cells in the large volume reservoir. B) Average monoclonality over four plates for each of 5 experiments illustrating consistency between experiments. The percentage of monoclonal wells approaches the expected distribution from limiting dilution.

While limiting dilution cannot achieve the efficiency of a cell sorting approach to monclonality, automating this plating and analysis as part of a larger automated cell line development system can reduce the effort needed to achieve the throughput required to identify optimal clones while also removing the need for additional costly tools.

References
Coller, H. A. & Coller, B. S. (1986) Poisson statistical analysis of repetitive subcloning by the limiting dilution technique as a way of assessing hybridoma monoclonality. Methods Enzymol, 121, 412-417. doi:10.1016/0076-6879(86)21039-3

Biomek Automated Workstations are not intended or validated for use in the diagnosis of disease or other conditions. Beckman Coulter Life Sciences genomic reagent kits are for research use only.

Helpful Links

Talk To An Expert