ChannelsStandard Sizes 2 channels: 0.5, 5.0 μm
Flow Rate 2.83 L per minute
Flow Control Critical orifice (requires 18” Hg vacuum, minimum 450 millibars)
Coincidence Loss 5% at 70,630,000 particles/m3
Particle Size Analysis Range 0.5 to 5.00 / 0.5 to 5
Throughput 2.83 L per minute
Operating Mode(s) Remote software-controlled sampling
Software Integrates with various software platforms for facility monitoring
Environmental Conditions Safe for indoor use only
Temperature Range Operating temperature range: 12°C to 29°C, Storage temperature range: -40°C to +70°C (ambient)
Humidity Restrictions 20% to 95% (non-condensing)
Power Requirements 6 VDC +/- 10%, 125 mA
Wiring, Fittings & Connectors Barb fitting for 0.32 cm ID inlet tubing, Barb fitting for 0.64 cm (ID) outlet tubing, DB-15
Material(s) 302 Stainless Steel
Weight 0.3 kg / 0.7 lb
Width 9.4 cm / 3.7 in
Depth 3.8 cm / 1.5 in
Height 6 cm / 2.4 in
Color Silver
Package Quantity 1

NOT ALL PRODUCTS ARE AVAILABLE IN ALL COUNTRIES
PRODUCT AVAILABILITY AND REGULATORY STATUS DEPENDS ON COUNTRY REGISTRATION PER APPLICABLE REGULATIONS

The listed regulatory status for products correspond to one of the below:
IVD: In Vitro Diagnostic Products. These products are labeled "For In Vitro Diagnostic Use."
ASR: Analyte Specific Reagents. These reagents are labeled "Analyte Specific Reagents. Analytical and performance characteristics are not established."
CE: Products intended for in vitro diagnostic use and conforming to European Directive (98/79/EC). (Note: Devices may be CE marked to other directives than (98/79/EC)
RUO: Research Use Only. These products are labeled "For Research Use Only. Not for use in diagnostic procedures."
LUO: Laboratory Use Only. These products are labeled "For Laboratory Use Only."
No Regulatory Status: Non-Medical Device or non-regulated articles. Not for use in diagnostic or therapeutic procedures.