With superior detection capabilities, the Navios provides accurate measurements for rare events critical to clinical diagnosis and monitoring. The flow cytometer clinical system delivers analytical excellence by coupling extraordinary sensitivity, resolution and dynamic range with high-speed data collection. The full-featured, easy-to-use software provides tools to improve the quality of results and optimize workflow in a clinical lab. 8 colors, 2 lasers (5 + 3 configuration).

Specifications

Platform Flow Analyser
Detection Photodiodes (forward & side scatter), Photomultiplier tubes (fluorescence)
Resolution HPCV is <2%
Performance <112 MESF for FITC, <78 MESF for PE, <15 MESF for PC5
Optics/Laser Power 488 nm solid state laser (22 mW), 638 nm solid state laser (25 mW)
Flow Cell BioSense 150 μm x 450 rectangular channel with an integral lens
Signal Processing Digital System with 4 decade data display
Software Navios Software
Power Requirements 50/60 Hz, 100-240 VAC
Temperature Range 15.5°C - 32°C
Humidity Restrictions 35%-85% (non-condensing)
Regulatory Status IVD, CE
Throughput 80 tubes/hour for 10,000 normal whole blood lymphocytes, Up to 88 tubes/hour at 10,000 events per second of concentrated beads
Installation Category II (per IEC 61010-1 standard)
Fluidics 11 L IsoFlow External Sheath Container, 20 L Waste Container, 1.5 L Flow Cleaning Fluid Tank , 1.5 L Internal Sheath Tank
Weight 104 kg / 229.3 lb
Width 95.3 cm / 37.5 in
Depth 70.1 cm / 27.6 in
Height 60.5 cm / 23.8 in
UNSPSC 41116015

NOT ALL PRODUCTS ARE AVAILABLE IN ALL COUNTRIES
PRODUCT AVAILABILITY AND REGULATORY STATUS DEPENDS ON COUNTRY REGISTRATION PER APPLICABLE REGULATIONS

The listed regulatory status for products correspond to one of the below:
IVD: In Vitro Diagnostic Products. These products are labeled "For In Vitro Diagnostic Use."
ASR: Analyte Specific Reagents. These reagents are labeled "Analyte Specific Reagents. Analytical and performance characteristics are not established."
CE: Products intended for in vitro diagnostic use and conforming to European Directive (98/79/EC). (Note: Devices may be CE marked to other directives than (98/79/EC)
RUO: Research Use Only. These products are labeled "For Research Use Only. Not for use in diagnostic procedures."
LUO: Laboratory Use Only. These products are labeled "For Laboratory Use Only."
No Regulatory Status: Non-Medical Device or non-regulated articles. Not for use in diagnostic or therapeutic procedures.