Enhance your workflow. Experience simplified high-throughput sample processing and more control with the JXN-26 three-phase centrifuge with powerful high speed features, low heat output, imbalance tolerance and quiet performance. Three-phase filtration divides materials into solids, liquids and gases. Has a maximum capacity of 6.0L and RPM of 26,000 (82,000 x g). The Friction Reduction System (FRS) means the machine runs quicker, quieter and cooler, with shorter cycle times with high-torque because of the exclusive Switched Reluctance (SR) drive. Stores up-to 1000 user-defined programs.

Specifications

Platform Avanti
Max Speed 26,000 rpm
Max g-Force 81,700.00 x g
BioSafe Yes
Acceleration Maximum or Timed ( 1-10 minutes from 0-500 rpm)
Deceleration Maximum, Timed ( 1-10 minutes from 500-0 rpm) or Coast
Speed Control ± 10 rpm of set speed or 0.1%, whichever is greater
Timed Run 99 min, 59 sec Continuous run, short run
Max Capacity Volume Range 1 x 4 L
Refrigeration System Non-CFC, non-ozone depleting refrigerant
Software Virtual Network Computing (VNC), Mobile Fuge app for Apple iOS and Android
Max Heat Dissipation 1.5 kW (5,120 BTU/hr)
Temperature Control ± 2 °C of set temperature after equilibration
Drive Type Cooling SR Drive* / Air cooled
Time (Actual Display) Real-time run monitoring, time elapsed or time remaining (hold run)
Temperature Range(s) From -20°C to 40°C in 1° increments
Rotor Chamber Diameter 51.3 cm / 20.2 in
Rotor ID Automatic
Zonal Capability Yes
Humidity Restrictions 80% (non-condensing)
Material(s) Steel, Structural Foam, Plastic
Finish Powder Coat Paint
Weight 310 kg / 683.4 lb
Width 71 cm / 28 in
Depth 86 cm / 33.9 in
Height 86 cm / 33.9 in
Clearances 7.6 cm (sides) & 16 cm (rear)
Pollution Degree 2
Max Noise Output 62 dBa
Power Requirements 200-240 VAC , 50/60 Hz

NOT ALL PRODUCTS ARE AVAILABLE IN ALL COUNTRIES
PRODUCT AVAILABILITY AND REGULATORY STATUS DEPENDS ON COUNTRY REGISTRATION PER APPLICABLE REGULATIONS

The listed regulatory status for products correspond to one of the below:
IVD: In Vitro Diagnostic Products. These products are labeled "For In Vitro Diagnostic Use."
ASR: Analyte Specific Reagents. These reagents are labeled "Analyte Specific Reagents. Analytical and performance characteristics are not established."
CE: Products intended for in vitro diagnostic use and conforming to European Directive (98/79/EC). (Note: Devices may be CE marked to other directives than (98/79/EC)
RUO: Research Use Only. These products are labeled "For Research Use Only. Not for use in diagnostic procedures."
LUO: Laboratory Use Only. These products are labeled "For Laboratory Use Only."
No Regulatory Status: Non-Medical Device or non-regulated articles. Not for use in diagnostic or therapeutic procedures.