Specifications

Platform Avanti
Max Speed 4,750 rpm
Max g-Force 5250 x g
Acceleration 10 acceleration rates, including maximum torque
Deceleration 11 deceleration rates, including maximum torque and no braking
Speed Control ± 10 increments from 200 - 10,200 rpm
Timed Run 1 minute-99 hours, 59 minutes or Continuous Hold
Max Capacity Volume Range 0 - 3 L
Refrigeration System Yes
Max Heat Dissipation 120 V: 4913 Btu/h (1.44 kW), 200-230V: 6551 Btu/h (1.92 kW)
Speed Display Actual rotor speed in 10 rpm increments or actual RCF in 10 x g increments
Temperature Control ±2 °C of Chamber temperature (after equilibration); applies to 4 to 25 °C temperature range
Time (Actual Display) Indicates run time remaining (HH:MM:SS)
User Defined Programs/Steps Access to 99 programs, 6 with one-touch
Temperature Range(s) From -10°C to +40°C in 1° increments
Installation Category II
Humidity Restrictions 80% (non-condensing)
Material(s) Painted Steel, Uncoated Plastic, Painted Aluminum, Plastic
Weight 120 kg / 264.6 lb
Width 75.6 cm / 29.8 in
Depth 70.3 cm / 27.7 in
Height 36.8 cm / 14.5 in
Clearances 7.6 cm: side & rear
Pollution Degree 2
Max Noise Output 58 dBa
Max Volume 3 L
Nominal Capacity 3 L
Power Requirements 200-230V, 8A, 50 Hz
UNSPSC 41103905

NOT ALL PRODUCTS ARE AVAILABLE IN ALL COUNTRIES
PRODUCT AVAILABILITY AND REGULATORY STATUS DEPENDS ON COUNTRY REGISTRATION PER APPLICABLE REGULATIONS

The listed regulatory status for products correspond to one of the below:
IVD: In Vitro Diagnostic Products. These products are labeled "For In Vitro Diagnostic Use."
ASR: Analyte Specific Reagents. These reagents are labeled "Analyte Specific Reagents. Analytical and performance characteristics are not established."
CE: Products intended for in vitro diagnostic use and conforming to European Directive (98/79/EC). (Note: Devices may be CE marked to other directives than (98/79/EC)
RUO: Research Use Only. These products are labeled "For Research Use Only. Not for use in diagnostic procedures."
LUO: Laboratory Use Only. These products are labeled "For Laboratory Use Only."
No Regulatory Status: Non-Medical Device or non-regulated articles. Not for use in diagnostic or therapeutic procedures.