The Optima AUC (A) An-60 Ti Bundle is the base analytical ultracentrifuge with 300 Hz, 10 µJM Xenon Flash Absorbance optics and also includes a 4-hole An-60 Ti analytical rotor, analytical cell, torque stand, counterbalance kit, accessory kit, and cell housing kit.

Specifications

Platform Optima AUC
Acceleration 5-400 rpm/second (above 500 rpm), 400 max acceleration
Deceleration Maximum, Slow & Off
Speed Control ±||actual rotor speed, ± 2 rpm at steady state
Timed Run 1 to 999 hrs, 59 mins (including hold runs)
Max Capacity Volume Range 0.00|0.45|mL
Operating System(s) LINUX
Software Experimental portal can be ran on MS IE 10 (or above); Safari 8 (or above); Chrome 42 (or above)
Speed Display 1,000 to rated speed in 100 rpm increments
Temperature Control 0.50|°C of set temperature after equilibration
Display 15.3" high-resolution color touch screen
Temperature Range(s) Ambient temperature range: 10 to 30°C, Set temp: 7-40° C (1°C increments at > 25-30°C ambient temp); 4-40°C (1°C increments at <25°C ambient temp)
Operating Mode(s) LINUX for remote monitoring; GUI (any OS)
Zonal Capability No
Humidity Restrictions <80% at <30°C (non-condensing)
Finish Rotor pad: Coated polycarbonate, Top & front surfaces: Polyurethane, Other surfaces: Kydex plastic & polyurethane painted aluminum
Weight /
Width /
Depth /
Height /
Clearances 5.1 cm (sides) & 15.2 cm (rear)
Pollution Degree 2
Power Requirements 30 A branch circuit, 50/60 Hz, 200-240 VAC

NOT ALL PRODUCTS ARE AVAILABLE IN ALL COUNTRIES
PRODUCT AVAILABILITY AND REGULATORY STATUS DEPENDS ON COUNTRY REGISTRATION PER APPLICABLE REGULATIONS

The listed regulatory status for products correspond to one of the below:
IVD: In Vitro Diagnostic Products. These products are labeled "For In Vitro Diagnostic Use."
ASR: Analyte Specific Reagents. These reagents are labeled "Analyte Specific Reagents. Analytical and performance characteristics are not established."
CE: Products intended for in vitro diagnostic use and conforming to European Directive (98/79/EC). (Note: Devices may be CE marked to other directives than (98/79/EC)
RUO: Research Use Only. These products are labeled "For Research Use Only. Not for use in diagnostic procedures."
LUO: Laboratory Use Only. These products are labeled "For Laboratory Use Only."
No Regulatory Status: Non-Medical Device or non-regulated articles. Not for use in diagnostic or therapeutic procedures.