Product Specifications

Platform APC Instruments
Performance ISO 21501-4 unit-to-unit accuracy & reproducibility
Industry Standards ISO 21501-4 , ISO 14644-1:2015, FCC, CE
Light Source Long Life Laser™ Diode
Particle Size Analysis Range 0.5 - 10 μm
Flow Rate 28.32 L per minute
Coincidence Loss 10% at 20,000,000 particles/m³
Sample Volume 10 µm
Throughput 28.32 L per minute
Interfaces & Protocols RS-485, RS232, Modbus RTU Communication Protocol | FXB Communication Protocol
Data Storage 1,000 samples (oldest records overwritten when buffer is full)
Threshold Setting 0.5 µm
Zero Count ISO 21501-4 & JIS B 9921 (1 count or less in 5 min; 95% UCL)
Count Alarms, Display & Cycles User-defined alarm counts displayed via LED on box or remote LED
Environmental Conditions Indoor Environment
Temperature Range Operating temperature range: 10°C to 32°C, Storage temperature range: -40 to +70°C
Humidity Restrictions 80% (non-condensing)
Power Requirements 24 VDC +/- 10% | 60 W maximum
Software Integrates with various software platforms for facility monitoring
Operating Modes Automated or remote sampling
Autoclavable No
Nonpyrogenic Yes
Weight 4.6 kg(10.1 lb)
Width 17.2 cm(6.8 in)
Depth 31.8 cm(12.5 in)
Height 11.5 cm(4.5 in)
Color Silver
Package Quantity 1