Neil Tiedeman, Beckman Coulter, Inc.Description:This information bulletin describes how to use the maintenance programs supplied with the ELx405 integration. Purpose:The ELx405 Microplate washer provides five maintenance programs that are stored on the washer and run through the maintenance selection on the washer keypad. When the washer is integrated onto a Biomek, it is difficult to access the keypad and view the 2-line LCD display. The ELx405 integration includes five Biomek methods that execute the maintenance programs, so that it is not necessary to access the keypad. Methods:Five Biomek methods are imported into the current Biomek project when the washer is installed. For Biomek Software version 2.5, they are stored in the Biomek methods directory. Each Biomek method runs one or more wash protocols defined using the ELx405 SILAS module. A description of each method follows. Note: The maintenance methods include an Instrument Setup step with a microplate in the washer position. However, these methods do not require a plate in the washer position. The Instrument setup step is used to prevent the Biomek method from asking that a plate be placed at the washer.
Example:This example demonstrates how to edit the maintenance program settings. It will guide you through how to change the valve port assignment for the DAY_RINSE program. Note: Once a valve port selection has been changed for a given protocol, the change will be retained for future use. Therefore, it is important to verify proper valve port configuration each time maintenance is performed on the ELx405 washer. Note: The DECONTAMINATION and LONG_SHUTDOWN methods utilize multiple ELx405 protocols; therefore, it is important to change valve ports for all protocols when changing valve port assignments for these methods. To change a valve port assignment: 1. From the Instrument menu, select Device Editor. Device Editor appears (Figure 1).
2. In Device, select ELx405 and choose Action Commands. ELx405 appears (Figure 2).
3. In Action, choose Run User Defined Program, and select the desired ELx405 protocol (BC_DAY_RINSE), and choose Edit. Bio-tek Autowasher Control Properties appears (Figure 3).
4. In Buffer Valve, select the desired valve port for the selected ELx405 maintenance protocol from the drop-down list and choose OK. The maintenance protocol is updated to use the newly selected valve port. 5. Repeat steps 2-4 for all ELx405 protocols used in the pre-configured Biomek method to be executed. AUTOCLEAN Program:The latest release of the ELx405 SILAS module (November, 2005) offers automation of the AUTOCLEAN program for washers with a built-in ultrasonic cleaner. It is expected that this program would be run at the end of the day after all washing is complete. There will be a program called AUTOCLEAN in the list of available wash protocols. This protocol consists of one or more cleaning loops. Multiple loops are used when cleaning with more than one type of fluid. The AUTOCLEAN program can be edited through the ELx405 SILAS module. To edit the AUTOCLEAN program: 1. From the Instrument menu, select Device Editor. Device Editor appears (Figure 4).
2. In Device, select ELx405 and choose Action Commands. ELx405 appears (Figure 5).
3. In Action, choose Run User Defined Program, and select the AUTOCLEAN program. Choose Edit. Autoclean Properties appears (Figure 6).
4. Choose Add to add a new loop to the protocol. Choose Modify to edit an existing loop. Choose OK to save your changes, or Cancel to exit without saving.
5. Choose OK to close the ELx405 dialog, and choose Configure Options. ConfigDialog appears (Figure 8).
6. Check the box Run AutoClean program after method finishes and choose OK. The AUTOCLEAN program shall run automatically at the end of any SAMI® EX, SAMI NT, or Biomek method that runs the ELx405 washer. Note: It is NOT necessary to specify a time estimate for the AUTOCLEAN program, unless you plan to execute it explicitly using a Biomek step or SAMI node. It shall run automatically at the end of a SAMI EX, SAMI NT, or Biomek method. For more information about the AUTOCLEAN program, see Chapter 4: Preventive Maintenance in the ELx405 Microplate Washer Operator’s Guide. Installation and Version requirements:1. Biomek FX version 2.5 software or later. * All trademarks are the property of their respective owners.