eLabNotebook > Nucleic Acid Prep & Purification > Plasmid DNA Purification > Promega SV 96 Two-Plate Biomek® FX


A Wizard Approach to Generating High Quality Plasmid DNA on the Biomek® FX Using the WIZARD* SV96 Plasmid DNA Purification System from Beckman Coulter

Monique Mason, John Snider, Graham J. Threadgill, Ph.D., and Laura Pajak, Ph.D. 
Beckman Coulter, Inc.

For additional information try our Maintenance and Troubleshooting section or post a question to our Biomek User Forum.  New members can subscribe to our User Forum here.

Problem Possible Cause Solution
Low or no yield:
  • Poor cell lysis: Over grown bacterial culture resulting in incomplete resuspension of bacterial cell pellet or clogging of Lysate Clearing plate.
  • Limit cell density (as measured at A600) of 1.0 to 4.0 OD600 per well.
  • Overgrowth of bacterial culture by nontransformed cells or bacterial culture is too old.
  • Inoculate antibiotic-containing media with single bacterial colony from a fresh plate. Make sure antibiotic is present in the media.
  • No ethanol or not enough ethanol added to wash solution.
  • Prepare the wash solution according to the instructions provided along with the kit.
  • Inaccurate quantitation of plasmid DNA yield.
  • Run the sample on an agarose gel along with known quantity of DNA ladder, stain the gel with ethidium bromide and quantitate by comparing the intensities.
  • Incomplete cell lysis due to poor resuspension of bacterial pellets.
  • Thoroughly resuspend cell pellet. Recommended centrifuge speed is 1,500 x g. Excessive centrifugation will result in poor resuspension.
  • Poor resuspension of bacterial pellets.
  • Use the full resuspension cycle included in the Wizard. Adjust the volume of the resuspension solution to the default value (250 µl) or greater for large bacterial pellets.
  • Over growth of bacterial cultures, particularly with strains containing Endonuclease I (EndA+) such as MC1061, JM101, BL21.
  • Grow cultures for a shorter time period, especially when using TB broth (16 hrs). LB media 16-24 hrs. Saturation conditions may cause cell lysis during culture resulting in loss and degradation of plasmid DNA.
  • Low vacuum pressure.
  • Adjust vacuum pressure to 15” Hg with the vacuum valve closed; reopen the valve before processing bacterial plates.
Filter plates not releasing from manifold gaskets
  • Worn, older gaskets.
  • Replace gaskets on manifold collar and spacer.
  • Vacuum pressure too high.
  • Adjust vacuum pressure to 15” Hg; alternatively, increase the time in the “pause” steps after the vacuum steps in the method to provide more time to allow for the system to equilibrate to atmospheric pressure.
Genomic DNA contamination
  • Too many mix cycles at resuspension step.
  • Reduce number of mix cycles at resuspension step.
Poor or no restriction digestion
  • Ethanol contamination.
  • Prevent ethanol contamination by allowing full 10 min of drying and blotting of the filter plate.
  • Too much DNA used in restriction digestion mix.
  • Quantitate DNA accurately by agarose gel/ethidium bromide electrophoresis.
  • Concentration of restriction enzyme and length of digestion is not optimal.
  • Check the manufacture’s suggestions for restriction enzymes including recommended restriction buffer, concentrations, and suggested temperature. This is particularly important when using multiple enzymes in a single digest reaction.
A ratio of A260/A280 that is less than 1.7
  • Protein contamination.
  • Overgrown cultures may result in carryover of protein by overloading the SV 96 system. Repurify plasmid DNA.
  • Spectrophotometric reading performed using water (<pH 6).
  • Perform spec reading in neutral water or TE (pH 7.0- 8.0). Some purified waters are acidic and cause a increase in the A 280 reading, making it appear like protein contamination.
Poor or no results with automated fluorescent DNA sequencing
  • Ethanol contamination.
  • Make sure there is no ethanol contamination by allowing the full 10 min drying and affixing fresh blot paper for the blotting step.
  • Too much or too little DNA used in sequencing reaction.
  • Quantitate DNA accurately by agarose gel/ethidium bromide electrophoresis in addition to spectrophotometric analysis.
  • TE buffer used for plasmid DNA elution.
  • Elute plasmid DNA only in nuclease free water provided in the kit.
Reagents are not delivered appropriately
  • Incorrect positioning of reagents when running method.
  • Consult the instrument setup screen to ensure reagents are placed in the appropriate location.
Crashes occur when moving labware with Gripper
  • Gripper fingers are worn, torn, or bent.
  • Replace worn or torn finger pads, and bent fingers as needed.
  • The gripper or worksurface is inaccurately framed.
  • Reframe the deck as necessary. Check framing by moving a piece of labware on the deck before running the method.
  • Labware placed in the wrong position.
  • Check instrument set up and make sure all labware is in correct location.
  • Improper labware definition.
  • Review definition in the Labware Editor and update if required. The labware can be reimported using the Import file.
Cannot move collar from the SPE ALP to the Holder
  • Gripper inaccurately framed.
  • Reframe gripper and/or adjust the squeeze/unsqueeze values for the collar and collar spacer in the labware editor.
  • Wrong 1x1 ALP placed adjacent to the Holder ALP.
  • Place the 1x1 with the cut-out sides adjacent to the Holder ALP. The cut-away sides allow the grippers to open to pick up/release the collar on the Holder ALP.
Vacuum bottle collapsing
  • Source vacuum is too high.
  • Check that the dead end vacuum pressure is 15”Hg.
  • Inappropriate bottle material.
  • Check base of bottle for “PP” (Polypropylene) designation. Replace with bottle made of correct material.

* All trademarks are the property of their respective owners. Where applicable, the PCR process is covered by patents owned by Roche Molecular Systems, Inc., and F. Hoffman-LaRoche, Ltd.

eLabNotebook Sitemap

©1998 - 2006 Beckman Coulter, Inc.