Millipore's Fully Automated Method for PCR Purification Using Montáge*
PCR96 Cleanup Kits on the Biomek® FX Nucleic Acid Preparation
System
Millipore– www.millipore.com
Configuration of the Biomek FX® Worksurface
(Click to Enlarge)
Prior to starting:
Make sure deck is set up as follows:
| SPE1: | Assembled Millipore manifold |
| P7: | P200 Tips |
| P8: | V-bottom collection plate for storage |
| P9: | MultiScreen96-PCR? plate |
| P10: | Reservoir for tip washing |
| P11: | Reservoir w/resuspension solution |
| P12: | Thermocycler plate with PCR samples |
Procedure
| 1. | Transfer 100µl of unpurified PCR product from thermocycler sample plate (P12) to
MultiScreen96-PCR plate (P9). Wash tips before unloading.
|
| 2. | Move MultiScreen96-PCR to Vacuum Manifold at SPE1.
|
| 3. | Apply vacuum for 10 minutes.
|
| 4. | Move MultiScreen96-PCR back to position P9.
|
| 5. | Load tips to add 100µl resuspension solution from reservoir on P11 to MultiScreen96-PCR plate (P9).
|
| 6. | Mix each sample 25 times at 100% speed and recover to collection plate (P8).
|
Figure 1. Agarose gel of purified PCR products
Figure 1. Representative samples from PCR purification using a Millipore Montage PCR96
Cleanup Kit on the Beckman Biomek FX. A 657 base pair product was amplified in a 100µl
PCR reaction. Samples were filtered through a MultiScreen96-PCR plate for 10 minutes and
recovered in 100µl of resuspension solution by mixing. M=Low DNA Mass ladder (8µl, Gibco
Life Technologies). 10µl of purified PCR samples were loaded into each well of a 1% agarose
gel. U=Unpurified PCR reaction (10µl).
|
Table 1. Average product recovery (%)
| 657 base pair product |
| Mean | 110% |
| Stdev | 21% |
| CV | 19% |
| Min | 66% |
| Max | 136% |
Table 1. Recovery of purified PCR products seen in Figure 1. 100µl PCR reactions
(657bp) were purified on the MultiScreen96-PCR plates from the Millipore Montage
PCR Cleanup Kit. Samples were resuspended in 100µl of TE buffer. Percent recovery
was determined by gel electrophoresis, ethidium bromide staining and densitometry.
Error rates for this analysis method can be 10-15%.
|
|
Conclusion:
Using the Millipore Montage PCR96 Cleanup Kit allows for removal of excess primers
and salts from the starting PCR reaction mix. Recoveries for a 657bp PCR product are
90% or greater for the conditions outlined in this procedure. The purified products may
be used for downstream applications such as sequencing, microarrays and restriction
digest without further purification. The process is simple, fast and automation
compatible.
|