Beckman Coulter, Inc.
 

Products

   
   
 
 

Optima™ TLX Personal Benchtop Ultracentrifuges--
Specifications and Ordering Information

Printer Friendly


Related Product Information

Specifications | Ordering Information

Specifications

  Optima TLX (627,000 x g; 120,000 rpm)
Speed control ± 50 rpm of set speed
Sample imbalance tolerance Up to 10% of volume in opposite tubes
Drive cooling* Air-cooled*
Refrigeration system* Solid state, no CFCs/ODCs*
Set temperature range 2°C to 40°C in 1°C increments
Temperature control ± 2°C of set
Ambient operating range 15°C to 35°C
User-settable programs 10 user programs; all user programs have up to five steps
Acceleration/Deceleration rates 10 Accel/11 Decel
Moisture-purging vacuum system Yes
Heat output 0.6 kW/hr (1,750 BTU/hr)
Sound level 3 ft (1 m) away < 60 dB
Remote control via RS-232 Optional
HEPA filter Optional
Dimensions H x W x D in (mm) 21 x 29 x 23 (533 x 737 x 584)
Weight 178 lb (81 kg)
* Watercooling and internal cooling units are not required.

back to top

Ordering Information

  Maximum
Force
Maximum
Capacity
220/240 VAC 50 Hz 110 VAC
60 Hz
100 VAC
50/60 Hz
Optima TLX 627,000 x g 40.8 mL
(8 x 5.1 mL)
361544 361545 361546

Optima TLX ultracentrifuges are UL/CUL approved and bear the CE marking.

 

MLA-130 Rotor MLA-80 Rotor MLN-80 Rotor MLS-50 Rotor
MLA-130
1,019,000 x g
20 mL (10 x 2 mL)
MLA-80
444,000 x g
64 mL (8 x 8 mL)
MLN-80
390,000 x g
64 mL (8 x 8 mL)
MLS-50
268,000 x g
20 mL (4 x 5 mL)

back to top