Cell Lab Quanta SC Flow Cytometer |
| Absolute Count Performance* |
Accuracy: ±5%
Reproducibility (CV): < 5%
Concentration: 3 x 104 to 2 x 106
|
| Fluorescence Performance Specifications |
Data acquired using Digital Signal processing and displayed on user selectable
linear or 4 decade logarithmic scales.
Resolution: < 3.0 %
HPCV (using TRBC Calibration Standard stained with DAPI and utilizing the mercury arc lamp)
< 3.0 % HPCV (using Flow-Check™ Fluorospheres utilizing the 488 nm diode laser) |
| Sizing Performance |
Electronic impedance with a measurement
range of 3.0 to 40 microns diameter with the standard 125 µm
flow cell. |
| Optics Performance |
Excitation:
Mercury arc excitation optimized at 366, 405, and 435
488 nm laser diode excitation, 2-22 mW user adjustable
Fluorescence Detectors:
3 broad range ultra sensitive photomultiplier tubes
Optical Alignment:
Computer assisted and controlled
Optical Coupling:
1.25NA oil immersion objective
Emission Filters:
Standard Set 460 BP, 525BP, 575BP and 670 LP |
| Parameters |
Electronic Volume, side scatter, time, and 3 color
detection with log/linear and FC/FSD options |
| Software |
Cell Lab Quanta SC Software for Instrument Control,
Data Acquisition and Data Analysis |
| Fluidics Specifications |
Flow Cell Patented 125 µm triangular flow cell (standard)
Sample rate: 4.17µL to 100 µL per minute for 125 µm flow cell
Delivery system: Vacuum pump/motorized syringe pump
Minimum sample size: 150 µL
Maximum sample size: 2 mL |
| Installation Requirements |
Power: 100 V, 120 V, or 220 V, 50/60 Hz
Wattage consumed: 500 watts (total system)
Operating temperature: 60°F to 85°F (16°C to 29°C)
Physical dimensions:
22 W x 26 L x 18 H (in.)
55.9 W x 67.3 L x 45.7 H (cm.)
Instrument weight: 77 lbs (34.9 kg) |
*These specifications were validated with an internal 10 µm bead that had an assigned count value. Your results may vary depending on your application and/or sample type.