Specifications

 
PARADIGM
General
Light Source High Powered LED or Flash Lamp
Plate Formats 6- to 1536-well
Temperature Range 4°C above ambient to 45°C
 
Uniformity ± 0.75°C
 
Shaking Linear, Orbital
Endpoint Measurement All Modes
 
Kinetic Measurement All Modes
Well Area Scanning All Modes
Wavelength Scanning Photometry
Dimensions W ×H ×D cm (in.) 39 (15.4) ×45 (17.7) ×63 (24.8)
Weight, without Cartridges 45 kg (99 lbs)
Power Source 100-240 VAC, 2 A, 50/60 Hz
Robotic Compatible YES

Typical Performance

Fluorometry (Fluorescein)
Stop-Read
On-The-Fly
384-well (75 μL)
 
0.100 fmol (2 min)
0.200 fmol (1 min)
 
1536-well (8 μL)
 
0.030 fmol (4 min)
0.050 fmol (1.5 min)
 
Fluorescence Polarization (Fluorescein 1 nM, SD)
384-well (75 μL)
 
3 mP (2 min)
6 mP (1 min)
1536-well (8 μL)
 
6 mP (4 min)
12 mP (1.5 min)
Time Resolved Fluorescence (Europium)
384-well (75 μL)
 
5 amol (2 min)
8 amol ( 1min)
 
1536-well (8 μL)
 
1.5 amol (4 min)
3 amol (1.5 min)
 
Luminescence (ATP)
 

 
384-well (75 μL)
0.5 fmol (3 min)
 
Photometric
Wavelength Range
 
230-1000 nm
 
Wavelength Selection
 
Monochromator (1.0 nm increments)
 
Wavelength Bandwidth
 
4.0 nm
 
Wavelength Accuracy
 
± 1 nm
 
Wavelength Repeatability
± 0.5 nm
Photometric Range
 
0 – 3.5 OD
Photometric Resolution
 
0.0001 OD
Photometric Accuracy
 
± (2% + 0.010 OD), 0-2 OD @ 405 nm
Photometric Precision
± (1% + 0.005 OD), 0-2 OD @ 405 nm

 

Ordering Information 

Part No. Description
A41572 Basic Configuration – Base Instrument; Detection Cartridges: Absorbance, Multimode, Fluorescence Intensity (fluorescein - rhodamine), Fluorescence Polarization (fluorescein), Cisbio HTRF; Multimode Analysis Software
A41570 Premiere Configuration – Base Instrument; Detection Cartridges: Absorbance, Multimode, Fluorescence Intensity (fluorescein - rhodamine), Fluorescence Intensity (coumarin - fluorescein), Fluorescence Polarization (fluorescein), Fluorescence Polarization (rhodamine), Time Resolved Fluorescence; Luminescence; Cisbio HTRF; Multimode Analysis Software
A41573 HTRF Configuration – Base Instrument; Cisbio HRTF Detection Cartridge; Multimode Analysis Software
 



 

 
 