| Available
Parameters |
| WBC |
MO% |
Hgb |
Retic # |
*MSCV
(Mean Sphered Cell Volume) |
| NE# |
EO# |
HCT |
Retic % |
Plt |
| NE%
|
EO%
|
MCV |
IRF (Immature Reticulocyte Fraction)
|
MPV (Mean Platelet Volume)
|
|
LY# |
BA# |
MCH |
MRV (Mean Reticulocyte
Volume) |
*PCT (Platelecrit) |
| LY%
|
BA% |
MCHC |
*HLS#
(High Light Scatter reticulocyte number)
|
*PDW
(Platelet Distribution Width)
|
|
MO# |
RBC |
RDW |
*HLS%
(High Light Scatter reticulocyte percent) |
|
| LH
750 Throughout |
LH755
Throughput |
Sample
Requirements |
| Mode |
Sample/Hr |
Mode |
Sample/Hr |
|
| CBC
Only |
110 |
CBC
Only |
105 |
Max.
300 µL Automatic Mode |
| CBC/Diff |
110 |
CBC/Diff |
100 |
Max.
550 µL Automatic Mode w/SlideMaker |
| CBC/Diff/Retic |
45 |
CBC/Diff/Retic |
45 |
Max.
200 µL Manual Mode |
|
LH
750 Workstation |
| Minimum
Specifications |
256MB
RAM |
17”
Flat Screen Monitor |
| Microsoft®
Windows 2000 Operating System |
20GB
Hard Drive |
Tower
configuration computer |
| Preemptive
Multitasking |
CD-ROM/DVD
Player |
Optical
Mouse |
| 1GHz
Processor |
Data/Fax
Modem |
|
Improved Linearity/Operating
Ratings
| Parameters |
Linearity
Range |
Operating
Range |
| WBC |
0.00 to 400.00
x 103 cells/µL |
0.00 to 900.00
x 103 cells/µL |
| RBC |
0.00 to 8.00
x 106 cells/µL |
0.00 to 20.00
x 106 cells/µL |
| Hgb |
0.0 to 25.0
g/dL |
0.0 to 99.9
g/dL |
| Plt |
0 to 3000 x
103 cells/µL |
0 to 5000 x
103 cells/µL |
| NE#, LY# |
0 to 100 x
103 cells/µL |
0 to 500 x
103 cells/µL |
| MO#, EO#, |
|
|
| BA# |
|
|
| RET% |
0 to 30% |
0.0 to 100.0% |
| RET# |
0.0 to 0.750
x 106 cells/µL |
0.0 to 999.9
x 106 cells/µL |
Precision
| Parameters |
Limits |
Parameters |
Limits |
| WBC
at 9 to 11 x 103 cells/µL |
£
1.7% CV |
NE%
at 50 to 60% |
2SD
£
3.0 |
| RBC
at 4.5 to 5.5 x 106 cells/µL |
£
0.8% CV |
LY%
at 25 to 35% |
2SD
£
3.0 |
| Hgb
at 14 to 16 g/dL |
£
0.8% CV |
MO%
at 5 to 10% |
2SD
£
2.0 |
| MCV
at 80 to 90 fL |
£
0.8% CV |
EO%
at 2 to 5% |
2SD
£
1.0 |
| RDW
at 12 to 14% |
£
2.2% CV |
BA%
at 0.5 to 1.5% |
2SD
£
1.0 |
| Plt
at 280 to 320 x 103 cells/µL |
£
3.3% CV |
RET%
at 0.0 to 0.49% |
16.5%
CV or 1SD £
0.23 |
| Plt
at 90 to 110 x 103 cells/µL |
£
6.6% CV |
RET%
at 0.5 to 1.49% |
14.5%
CV or 1SD £
0.23 |
| Plt
at 10.0 to 15.0 x 103 cells/µL |
£
14.0% CV |
RET%
at 1.5 to 4.0% |
11.0%
CV or 1SD £
0.68 |
| MPV
at 8 to 10 fL |
£
2.2% CV |
RET%
at 4.01 to 15.0% |
5.5%
CV or 1SD £
0.68 |
Dimensions
| Component |
Height |
Width |
Depth |
Weight |
| Analyzer |
35
in |
40
in |
24
in |
205
lbs |
| |
88.9
cm |
101.6
cm |
61
cm |
93.2
Kg |
| Diluter |
20
in |
39
in |
22
in |
120
lbs |
| |
50.8
cm |
99.1
cm |
55.9
cm |
54.4
Kg |
| Power
Supply |
25
in |
15
in |
26
in |
125
lbs |
| |
63.5
cm |
38.1
cm |
66.0
cm |
56.7
Kg |
| Slide
Maker |
19.75
in |
29.5
in |
27
in |
90
lbs |
| |
50.2
cm |
74.9
cm |
68.6
cm |
40.9
Kg |
| Slide
Stainer |
19.75
in |
23.75
in |
30
in |
80
lbs |
| |
50.2
cm |
60.3
cm |
76.2
cm |
36.4
Kg |
*For
research use only. Not for use in diagnostic procedures.
Microsoft®
Windows®
and Windows®
NT are trademarks of Microsoft Corporation.
NT is a trademark of Northern Telecom,
Inc.

|