| Available Parameters |
| WBC |
MO% |
Hgb |
Retic # |
*MSCV (Mean Sphered Cell Volume) |
| NE# |
EO# |
HCT |
Retic % |
Plt |
| NE%
|
EO%
|
MCV
|
IRF (Immature Reticulocyte Fraction)
|
MPV (Mean Platelet Volume)
|
|
LY# |
BA# |
MCH |
MRV (Mean Reticulocyte Volume) |
*PCT
(Platelecrit) |
| LY%
|
BA%
|
MCHC
|
*HLS# (High Light Scatter reticulocyte number)
|
*PDW (Platelet
Distribution Width)
|
|
MO# |
RBC |
RDW |
*HLS% (High Light Scatter reticulocyte percent) |
|
|
|
| LH
750 Throughout |
LH755
Throughput |
Sample
Requirements |
| Mode |
Sample/Hr |
Mode |
Sample/Hr |
|
| CBC
Only |
110 |
CBC
Only |
105 |
Max. 300 µL
Automatic Mode |
| CBC/Diff |
110 |
CBC/Diff |
100 |
Max. 550 µL
Automatic Mode w/SlideMaker |
| CBC/Diff/Retic |
45 |
CBC/Diff/Retic |
45 |
Max. 200 µL
Manual Mode |
|
|
|
LH 750 Workstation |
| Minimum
Specifications |
256MB
RAM |
17"
Flat Screen Monitor |
| Microsoft®
Windows 2000 Operating System |
20GB
Hard Drive |
Tower
configuration computer |
| Preemptive
Multitasking |
CD-ROM/DVD
Player |
Optical
Mouse |
| 1GHz
Processor |
Data/Fax
Modem |
|
|
|
Improved Linearity/Operating Ratings
| Parameters |
Linearity Range |
Operating Range |
| WBC |
0.00
to 400.00 x 103 cells/µL |
0.00
to 900.00 x 103 cells/µL |
| RBC |
0.00
to 8.00 x 106 cells/µL |
0.00
to 20.00 x 106 cells/µL |
| Hgb |
0.0
to 25.0 g/dL |
0.0
to 99.9 g/dL |
| Plt |
0
to 3000 x 103 cells/µL |
0
to 5000 x 103 cells/µL |
| NE#,
LY# |
0
to 100 x 103 cells/µL |
0
to 500 x 103 cells/µL |
| MO#,
EO#, |
|
|
| BA# |
|
|
| RET% |
0
to 30% |
0.0 to 100.0% |
| RET# |
0.0
to 0.750 x 106 cells/µL |
0.0
to 999.9 x 106 cells/µL |
|
|
|
Precision
| Parameters |
Limits |
Parameters |
Limits |
| WBC
at 9 to 11 x 103 cells/µL |
£
1.7%
CV |
NE%
at 50 to 60% |
2SD £
3.0 |
| RBC
at 4.5 to 5.5 x 106 cells/µL |
£
0.8%
CV |
LY%
at 25 to 35% |
2SD £
3.0 |
| Hgb
at 14 to 16 g/dL |
£
0.8%
CV |
MO%
at 5 to 10% |
2SD £
2.0 |
| MCV
at 80 to 90 fL |
£
0.8%
CV |
EO%
at 2 to 5% |
2SD £
1.0 |
| RDW
at 12 to 14% |
£
2.2%
CV |
BA%
at 0.5 to 1.5% |
2SD £
1.0 |
| Plt
at 280 to 320 x 103 cells/µL |
£
3.3%
CV |
RET%
at 0.0 to 0.49% |
16.5%
CV or 1SD £
0.23 |
| Plt
at 90 to 110 x 103 cells/µL |
£
6.6%
CV |
RET%
at 0.5 to 1.49% |
14.5%
CV or 1SD £
0.23 |
| Plt
at 10.0 to 15.0 x 103 cells/µL |
£
14.0%
CV |
RET%
at 1.5 to 4.0% |
11.0%
CV or 1SD £
0.68 |
| MPV
at 8 to 10 fL |
£
2.2%
CV |
RET%
at 4.01 to 15.0% |
5.5%
CV or 1SD £
0.68 |
|
|
|
Dimensions
|
Component
|
Height |
Width |
Depth |
Weight
|
| Analyzer |
35 in |
40 in |
24 in |
205 lbs |
| |
88.9 cm |
101.6 cm |
61 cm |
93.2 Kg |
| Diluter |
20 in |
39 in |
22 in |
120 lbs |
| |
50.8 cm |
99.1 cm |
55.9 cm |
54.4 Kg |
| Power Supply |
25 in |
15 in |
26 in |
125 lbs |
| |
63.5 cm |
38.1 cm |
66.0 cm |
56.7 Kg |
| Slide Maker |
19.75 in |
29.5 in |
27 in |
90 lbs |
| |
50.2 cm |
74.9 cm |
68.6 cm |
40.9 Kg |
| Slide Stainer |
19.75 in |
23.75 in |
30 in |
80 lbs |
| |
50.2 cm |
60.3 cm |
76.2 cm |
36.4 Kg |
|
|
*For
research use only. Not for use in diagnostic procedures.
Microsoft®
Windows® and Windows® NT are trademarks of Microsoft
Corporation.
NT
is a trademark of Northern Telecom, Inc.
.
|
|
|