Beckman Coulter, Inc.

Products

   
   
  Specifications
Printer Friendly


Related Product Information

Available Parameters
WBC MO% Hgb Retic # *MSCV (Mean Sphered Cell Volume)
NE# EO# HCT Retic % Plt
NE% EO% MCV IRF (Immature Reticulocyte Fraction) MPV (Mean Platelet Volume)  
LY# BA# MCH MRV (Mean Reticulocyte Volume) *PCT (Platelecrit)
LY% BA% MCHC *HLS# (High Light Scatter reticulocyte number) *PDW (Platelet Distribution Width)  
MO# RBC RDW *HLS% (High Light Scatter reticulocyte percent)  
LH 750 Throughout LH755 Throughput Sample Requirements
Mode Sample/Hr Mode Sample/Hr  
CBC Only 110 CBC Only 105 Max. 300 µL Automatic Mode
CBC/Diff 110 CBC/Diff 100 Max. 550 µL Automatic Mode w/SlideMaker
CBC/Diff/Retic 45 CBC/Diff/Retic 45 Max. 200 µL Manual Mode

LH 750 Workstation

Minimum Specifications 256MB RAM 17" Flat Screen Monitor
Microsoft® Windows 2000 Operating System 20GB Hard Drive Tower configuration computer
Preemptive Multitasking CD-ROM/DVD Player Optical Mouse
1GHz Processor Data/Fax Modem  
 

Improved Linearity/Operating Ratings

Parameters Linearity Range  Operating Range
WBC 0.00 to 400.00 x 103 cells/µL 0.00 to 900.00 x 103 cells/µL
RBC 0.00 to 8.00 x 106 cells/µL 0.00 to 20.00 x 106 cells/µL
Hgb 0.0 to 25.0 g/dL 0.0 to 99.9 g/dL
Plt 0 to 3000 x 103 cells/µL 0 to 5000 x 103 cells/µL
NE#, LY# 0 to 100 x 103 cells/µL 0 to 500 x 103 cells/µL
MO#, EO#,    
BA#    
RET% 0 to 30% 0.0 to 100.0%
RET# 0.0 to 0.750 x 106 cells/µL 0.0 to 999.9 x 106 cells/µL
 

Precision

Parameters Limits Parameters Limits
WBC at 9 to 11 x 103 cells/µL £ 1.7% CV NE% at 50 to 60% 2SD £ 3.0
RBC at 4.5 to 5.5 x 106 cells/µL £ 0.8% CV LY% at 25 to 35% 2SD £ 3.0
Hgb at 14 to 16 g/dL £ 0.8% CV MO% at 5 to 10% 2SD £ 2.0
MCV at 80 to 90 fL £ 0.8% CV EO% at 2 to 5% 2SD £ 1.0
RDW at 12 to 14% £ 2.2% CV BA% at 0.5 to 1.5% 2SD £ 1.0
Plt at 280 to 320 x 103 cells/µL £ 3.3% CV RET% at 0.0 to 0.49% 16.5% CV or 1SD £ 0.23
Plt at 90 to 110 x 103 cells/µL £ 6.6% CV RET% at 0.5 to 1.49% 14.5% CV or 1SD £ 0.23
Plt at 10.0 to 15.0 x 103 cells/µL £ 14.0% CV RET% at 1.5 to 4.0% 11.0% CV or 1SD £ 0.68
MPV at 8 to 10 fL £ 2.2% CV RET% at 4.01 to 15.0% 5.5% CV or 1SD £ 0.68

Dimensions

Component

Height Width Depth

Weight

Analyzer 35 in 40 in 24 in 205 lbs
  88.9 cm 101.6 cm 61 cm 93.2 Kg
Diluter 20 in 39 in 22 in 120 lbs
  50.8 cm 99.1 cm 55.9 cm 54.4 Kg
Power Supply 25 in 15 in 26 in 125 lbs
  63.5 cm 38.1 cm 66.0 cm 56.7 Kg
Slide Maker 19.75 in 29.5 in 27 in 90 lbs
  50.2 cm 74.9 cm 68.6 cm 40.9 Kg
Slide Stainer 19.75 in 23.75 in 30 in 80 lbs
  50.2 cm 60.3 cm 76.2 cm 36.4 Kg

*For research use only. Not for use in diagnostic procedures. Microsoft® Windows® and Windows® NT are trademarks of Microsoft Corporation. NT is a trademark of Northern Telecom, Inc.

.